Lab Dept:
|
Chemistry
|
Test Name:
|
VALPROIC ACID LEVEL
|
General Information
|
Lab Order Codes:
|
VALP
|
Synonyms:
|
Depakene; Depakane, Depakote; Dipropylacetic acid; VPA; Divalproex; Valproate
|
CPT Codes:
|
80164 – Dipropylacetic acid (valproic acid)
|
Test Includes:
|
Valproic acid concentration reported in ug/mL.
|
Logistics
|
Test Indications:
|
Valproic acid is used for treatment of seizures. Valproic acid may be used in conjunction with other antiepileptics against tonic-clonic and partial seizures. Peak levels are reached in 1-4 hours. The biologic half-life is about 15 hours.
|
Lab Testing Sections:
|
Chemistry
|
Phone Numbers:
Minneapolis:
Saint Paul:
|
|
612-813-6280
|
651-220-6550
|
Test Availability:
|
Daily, 24 hours
|
Turnaround Time:
|
4 hours
|
Special Instructions:
|
N/A
|
Specimen
|
Specimen Type:
|
Blood
|
Container:
|
Green top (Li Heparin) tube preferred
Alternate tube: Red-top (plain, no gel) tube
|
Draw Volume:
|
0.6 mL blood
|
Processed Volume:
|
0.2 mL plasma/serum
|
Collection:
|
Routine venipuncture. Mix tubes containing anticoagulant by gentle inversion.
|
Special Processing:
|
Lab Staff: Centrifuge specimen, remove serum/plasma aliquot into a plastic sample cup. Store at refrigerated temperatures.
|
Patient Preparation:
|
None
|
Sample Rejection:
|
Mislabeled or unlabeled specimen; specimens other than heparinized plasma or serum
|
Interpretive
|
Reference Range:
|
Therapeutic concentration: 50-100 ug/mL
|
Critical Values:
|
Toxic concentration: >200 ug/mL
|
Limitations:
|
N/A
|
Methodology:
|
Particle Enhanced Turbidimetric Inhibition Immunoassy (PETINIA)
|
References:
|
VALP Flex® reagent cartridge insert sheet, Siemens Healthcare Diagnostics Inc, 2008-02-29 717078.001 – US, Rev E
|
www.childrensmn.org/Lexi-Comp Copyright © 1978-2006 Lexi-Comp Inc
|
Clinical Significance, Dade Behring Inc., Glasgow Business Community, Mailbox 531, P.O. Box 6101, Newark, Delaware 19714
|
Jacobs & DeMott Laboratory Test Handbook (2001) Lexi-Comp, Inc, Hudson, OH, 5th Edition
|
Biorad Liquichek Unassayed Chemistry Control Product Insert, Bio-Rad Laboratories, Irvine, CA 92618
|
Updates:
|
5/1/2005: Critical value changed, previously listed as >150 ug/mL.
|