Clinics and Departments

Laboratory Services

Lab Dept:

Chemistry

Test Name:

IMATINIB LEVEL

General Information

Lab Order Codes:

IMAT

Synonyms:

Gleevec

CPT Codes:

No charge

Test Includes:

Imatinib (Gleevec) concentration measured in ng/mL

Logistics

Test Indications:

Useful to monitor therapeutic concentration in patients with Philadelphia chromosome–positive chronic myeloid leukemia (Ph+ CML). Monitoring may help in maximizing the drug's therapeutic effect, minimizing toxicity, and assesing patient adherence to oral therapy.

Lab Testing Sections:

Chemistry – Sendouts

Referred to:

Avantix Laboratories - CML & GIST Alliance Blood Level Testing Program

Phone Numbers:

Minneapolis:

Saint Paul:

 

612-813-6280

651-220-6550

Test Availability:

Daily, 24 hours

Turnaround Time:

2 – 7 days

Special Instructions:

Draw within 2 hr window of the next dose or anytime after 4 hr post dose. Special Tube Required. Call Children’s laboratories to obtain the Pink top tube to collect blood.

Specimen

Specimen Type:

Whole blood

Container:

Pink top (K2-EDTA) tube

Draw Volume:

6 mL blood

Processed Volume:

2 mL plasma

Collection:

Routine venipuncture

Special Processing:

Lab Staff: Centrifuge at 1500g for 10 minutes within 1hour of draw at room temp. Transfer 2 mL of K2-EDTA plasma into the polypropylene tube provided. Fill out the provided label & affix it to the tube with tape. Place the polypropylene tube contained K2-EDTA plasma inside the provided crush proof conical tube along with a cotton ball for absorbency. Place the conical tube in the corrugated box & seal the box w/ the provided biohazard sticker. Place the entire assembly in the FedEx Cinical Pak with the FedEx billable stamp outside to ship.

Store sample refrigerated and ship the same day at room temp or store at -20C until shipment. Sample should be shipped MON-THURS only.

Fill out & submit the test request form (provided in the kit) along with the specimen.

Patient Preparation:

None

Sample Rejection:

Specimens other K2-EDTA plasma; mislabeled or unlabeled specimens

Interpretive

Reference Range:

N/A

Critical Values:

N/A

Limitations:

N/A

Methodology:

High Performance Liquid Chromatography with Tandem Mass Spectrometry (LC-MS/MS)

References:

Avantix Laboratories - CML & GIST Alliance™ Blood Level Testing Program July 2009 1-866-990-0007



  Lab Test Directory
    Chemistry
    Coagulation
    Flow/Immunology
    Hematology
    Microbiology/ Virology
    Other Fluids
    Pathology
    Serology
    Transfusion Services
    Urine/Stool

  Search Lab Test Directory
    List All Documents

  Lab Staff

  E-mail Lab Services