Lab Dept:
|
Coagulation
|
Test Name:
|
FACTOR X (FAIRVIEW UNIVERSITY)
|
General Information
|
Lab Order Codes:
|
F10U
|
Synonyms:
|
Stuart Factor; Stuart-Prower Factor; Factor X Assay
|
CPT Codes:
|
85260 – Clotting; factor X (Stuart-Prower)
|
Test Includes:
|
Factor X level reported as a %.
|
Logistics
|
Test Indications:
|
Useful for the detection of a single factor congenital homozygous or heterozygous deficiency or an acquired deficiency due to a lack of Vitamin K, liver disease, or amyloidosis.
|
Lab Testing Section:
|
Coagulation - Sendouts
|
Referred to:
|
Fairview University Medical Center (Fairview Test: F10)
|
Phone Numbers:
Minneapolis:
Saint Paul:
|
|
612-813-6280
|
651-220-6550
|
Test Availability:
|
Daily, 24 hours
|
Turnaround Time:
|
4 hours if received on Monday – Friday: 0700-1730 or Saturday: 1000-1300. 1 day if received outside of these regular working hours. Stats are done within 1 hour of receipt in the laboratory.
For urgent testing after hours, contact Children’s laboratory and they will call Fairview’s tech on call.
|
Special Instructions:
|
Patient should not be receiving Coumadin® or heparin. If so, this should be noted on the request form. Heparin or warfarin therapy can affect certain coagulation factors or assays, preclude their performance, or cause spurious results.
Indicate when specimen is drawn from a line or a heparin lock. Deliver immediately to the laboratory.
|
Specimen
|
Specimen Type:
|
Whole blood
|
Container:
|
Light Blue top (Buffered Na Citrate 3.2%) tube
|
Draw Volume:
|
2.7 mL in a 3 mL tube (Minimum: 1.8 mL in a 2 mL tube)
Please respect fill line on tube.
|
Processed Volume:
|
Specimen will be processed at the reference lab facility.
|
Collection:
|
Routine venipuncture. Do Not use the first 2 mL’s of blood collected.
Patients with Hematocrit levels >55% must have a special tube made to adjust for the hematocrit, contact laboratory for special tube.
Fill tube completely. Mix thoroughly by gentle inversion.
|
Special Processing:
|
Lab Staff: Do Not centriguge. Send to reference lab in original Vacutainer® at room temperature.
Must arrive in reference lab within 24 hours of collection.
|
Patient Preparation:
|
Avoid heparin therapy for 2 days prior to test.
If the patient has a coagulation abnormality, apply direct pressure to the puncture site for 10 minutes; apply pressure bandage. Instruct the patient to leave the bandage on for 12 hours.
|
Sample Rejection:
|
Improper tube; clotted samples; underfilled tubes; overfilled tubes; mislabeled or unlabeled specimens
|
Interpretive
|
Reference Range:
|
Age:
|
Range (%):
|
<1 day:
|
22 – 72%
|
1 – 4 days:
|
26 – 54%
|
5 – 29 days:
|
34 – 64%
|
30 – 89 days:
|
45 – 73%
|
90 – 179 days:
|
53 – 89%
|
180 – 364 days:
|
58 – 98%
|
≥1 year:
|
60 – 140%
|
Critical Values:
|
N/A
|
Limitations:
|
Interpretation of results may be limited if patient is receiving anticoagulant therapy or if test is done more than 2 hours after collection.
|
Methodology:
|
PT clotting time correction of Factor 10 deficient plasma. Patient dilutions are compared to a known set of standard dilutions and a percentage is determined.
|
References:
|
Fairview University Diagnostic Laboratory Web Page June 2011
|
Updates:
|
10/7/2005: Reference ranges added for 0-5 days, previously only listed for greater than or equal to 1 year.
5/24/2010: Tubing of patient specimens is no longer prohibited.
6/15/2011: Reference range update
|