Clinics and Departments

Laboratory Services

Lab Dept:

Coagulation

Test Name:

TEG WITH PLATELET INHIBITION

General Information

Lab Order Codes:

TEGP

Synonyms:

Thromboelastrograph with platelet mapping; Platelet Inhibition; Platelet Mapping

CPT Codes:

85576 x5 – Platelet aggregation, each agent
85347 x2 – Coagulation time, activated
85384 x2 – Fibrinogen; activity
85390 x2– Fibrinolysins or coagulopathy screen, interpretation and report

Test Includes:

TEG parameters, AAPI (Platelet Arachidonic Acid) and AADP (Platelet Adenosine Diphosphate) reported as % Inhibition, TEG and/or TEG with heparin parameters (R, K, Angle, MA, G, LY30) will also be reported (if appropriate).

Logistics

Test Indications:

Assess hemorrhagic or thrombic risk by measuring rate of clot formation, strength and stability of clot; the effect of platelets, coagulation factors and cellular interactions, maximum platelet function.

Lab Testing Section:

Coagulation

Phone Numbers:

MIN Lab: 612-813-6280

STP Lab: 651-220-6550

Test Availability:

7 AM – 1330 PM, 7 days a week. Must be received in lab by 1330 PM.

Turnaround Time:

2 hours

Special Instructions:

Deliver immediately to the laboratory. Specimens MUST be walked to the laboratory and cannot be delivered via pneumatic tube. Must arrive in lab within 1 hour of collection. Indicate when specimen is drawn from a line or a heparin lock.

Specimen

Specimen Type:

Whole blood

Container:

Light Blue top (Buffered Na Citrate 3.2%) tube AND Green top (LiHep or NaHep)
NOTE: Both tubes MUST be drawn together

Draw Volume:

Blue top (Na Citrate) tube: 1.8 mL AND
Green top (LiHep or NaHep) tube: 3 mL (minimum)

*** Deliver BOTH TUBES immediately to the Lab.

Specimens must be hand delivered to the laboratory. No pneumatic tubing of TEG specimens.

Processed Volume:

Same as Draw Volume

Collection:

A clean venipuncture or line draw is essential. Mix specimen thoroughly by gentle inversion. Specimen must arrive within 1 hour of collection. Do not send via pneumatic tube system.

Patient’s with a hematocrit level >55% must have a special tube made to adjust for the hematocrit; contact laboratory for special tube.

Indicate when specimen is drawn from a line or heparin lock.

Special Processing:

Lab Staff: Do Not centrifuge. Do Not freeze. Deliver to TEG workstation for immediate testing.

Patient Preparation:

None

Sample Rejection:

Improper tube; clotted sample; underfilled tubes; overfilled tubes; mislabeled or unlabeled specimens

Interpretive

Reference Range:

TEG Test:

Range (all ages)

AAPI

No reference range (% Inhibition)

ADPI

No reference range (% Inhibition)

R

1.0 – 5.0 minutes

K

1.0 – 3.0 minutes

Angle

53.0 – 72.0 degrees

MA

50.0 – 70.0 mm

G

4.5 – 11.0 kd/sc

LY30

0.0 – 8.0 %

Critical Values:

N/A

Limitations:

N/A

Methodology:

TEG Viscoelastic Clot Assessment

References:

Haemonetics TEG 5000 Technical Manual (2008)

Updates:

8/6/2014: Updated Test Availability, previously listed as Daily, 24 hours.



  Lab Test Directory
    Chemistry
    Coagulation
    Flow/Immunology
    Hematology
    Microbiology/ Virology
    Other Fluids
    Pathology
    Serology
    Transfusion Services
    Urine/Stool

  Search Lab Test Directory
    List All Documents

  Lab Staff